ETF11 Electron Time-of-Flight Spectrometer

ETF11 Electron Time-of-Flight Spectrometer

Features:

  • Minimum electron energy ca. 20 eV (depending on the magnetic fields of the environment)
  • Time-of-flight resolution T/ΔT = ca. 100
  • Fieldfree drift tube with 350 mm standard length, encapsulated in a µ-metal shield
  • Drift tube region can be floated to positive or negative potentials:
    up to –180 V for decelerating fast electrons
    up to +180 V for accelerating slow electrons to reduce the effect of magnetic fields
  • Channelplate detector with two-stage chevron geometry, channelplates with 5 µm pore size for fast response times, 40 mm active diameter
  • Post-acceleration stage in front of the detector to get a uniform sensitivity for all electrons irrespective of their kinetic energy and to reduce the effects of magnetic fields penetrating through the back flange into the µ-metal shield
  • Conical entrance aperture with 3 mm diameter, the distance between the entrance aperture and the laser focus is also 3 mm. The distance between the tof connection flange and the laser focus is 354 mm (standard)
  • Only completely non-magnetic materials are used for the spectrometer, most of the parts including the drift tube and all screws are made from titanium
  • The electron beam entering through the entrance aperture is collimated in an extraction module, the acceptance angle (full angle) is ca. 45°. The only connection between the ionisation chamber and the tof chamber is the entrance aperture which allows to pump the tof spectrometer differentially
  • The vacuum chamber of the electron spectrometer is manufactured from non-magnetic stainless steel 1.4429 (equivalent to 316 LN). It consists of a 6” o.d. conflat (CF100) connection flange, a 6” o.d. conflat (CF100) pump flange (CF100) and a 2.75” o.d. conflat flange (CF38) for pressure metering
  • High voltage power supply in a 19” box, 3HE (132 mm) high, 350 mm deep

 Options:

  • Drift tube length other than standard length of 350 mm
  • Distance between tof connection flange and laser focus other than standard length of 354 mm
  • Accelerating/decelerating voltages different from standard values of +180V / -180 V, different maximal lens voltage

ETF11 Electron Time-of-Flight Spectrometer