Ion Multi Gauge Controller MG13
Description
As a big advantage of this instrument it can be possibleities to use up to four triggers, which can programmable for each channel separately, with own programmable hysteretic also as 'below' or 'above' options. The measurement channels are active immediately after applying the power, and ready to emision. Channel one is dedicated to control B-Agauge only. Really flexible settings of B-Agauge parameters allows to connect almost all of B-A gauges available on the market. Such a 2 channel controller combination is perfectly suitable in 2 stages pumps group units, where two ranges (low and high vacuum) has to be controlled. More over the automatic emission control facility can be used to automatic EMISSION of B-A gauge control related to the low vacuum measurement. All operating modes of the triggers and the transmitters are displayed and their status is indicated via the outputs.
The unit is also fully software controlled and web-enabled so that it can be operated over any network, including the internet.
Advantages
• 19” rack mounted and only 1U height
• Measurement range from atmospheric to 2x10-11 mbar
• Controls almost all commercially available Bayard-Alpert gauges
• Degassing of Bayard-Alpert with power and time limit
• Supports Ir/W/Thoria filaments
• Bayard-Alpert overpressure protection
• Selectable measurement units: mbar, Torr, Pascal
• Gas specific correction with one customized setting
• Universal power supply: 90-240VAC, 50/60Hz
• Unit over-temperature protection
• Audible alarm
• Four individually programmable set-points with threshold and hysteresis functions
• User-defined channel name
Applications
• Any high and ultra high vacuum systems
• Measurement of ultimate pressure (from atmosphereto UHV range)
• Bakeable and non bakeable systems
Technical data
Measurement chanels | 2 |
Display range | 2x10-11-2x105mbar |
Measurement range: Chanel 1: B-A gauge Channel 2: 0 - 10V DC input* |
2x10-11 mbar |
Display | LCD 2x16 characters 9,96mm high each line |
Setpoints ( individual programming for each channel possible ) |
3xNO, 1xNC |
Suitable Gauges in standard** | DI 200, DI 2000, TTR 211S, TTR91, PTR225, ITR90, ITR100, PTR 90, B-A gauge, MKS Baratron**** |
Hysteresis | adjustable |
Supply | 90÷240 VAC |
Operating temperature | 5-50°C |
Communication Interface*** (option) | RS232/485/422, USB, EtherNet/IP, Bluetooth |
Dimensions [mm]: (˝ of 1HE) | 240 x 260 x 44 (W x H x D) |
Weight, approx | 3,8 kg |
*related to the connected gauge
**any other can be implementedon request
*** Only one comms interface can be used at any one time and is chosen at time of order. However, this may be swapped/interchanged with any other comms interface at any future date via simple rear panel plug and play swap, without having to access the internals of the unit.
**** please refer to manual
Only one communication interface can be chosen at time of order (please refer to table) but this can be easily changed in the future via simple rear panel plugin connection modules, without ever having to access the internals of the unit.